Performance Analysis of Medical Syringe and Infusion Pumps: (Factorial Experiments Case Study on the Combined Effects of Syringe Size, Infusion Line Length and Tubing Compliance)

Document Type : Technical Note

Authors
1 Biomedical Engineering, Research Group, Research Center of Technology and Engineering, Standard Research Institute, Karaj, Iran
2 Department of Medical Physics and Engineering, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran
10.22034/asas.2026.588507.1113
Abstract
Medication errors are among the leading causes of serious adverse events in hospitalized patients, and a considerable number of deaths each year can be traced back to them. In syringe and infusion pumps, delivery accuracy is a safety-critical property, especially in intensive care and neonatal environments. Importantly, the accuracy of these devices is not dictated by the motor and drive mechanism alone. A group of external factors, namely the size and mechanical compliance of the mounted syringe, the length and internal diameter of the infusion line, and the material behavior of the tubing, exert a strong and frequently underestimated influence on performance.
This study sets out an integrated experimental and statistical investigation of the variables that shape syringe pump performance. Within a full factorial design, several analytical techniques were applied. The data set comprised more than 300 observations spanning three syringe sizes, three programmed flow rates, several tubing types, and two infusion line lengths.
They confirm that syringe size, tubing compliance, and line length are each statistically significant determinants of pump performance, with the largest effects appearing at the lowest flow rate (1.0 mL/h). Taken together, the findings show that larger syringe diameters magnify the effect of mechanical play in the drive train; that long infusion lines add both fluidic resistance and volumetric compliance, the latter being the dominant cause of delayed flow onset at low rates; and that compliant tubing stores elastic strain energy during start-up, producing a delayed onset of flow together with transient under-delivery.
Keywords
Subjects

[1]  Ali Tavakoli Golpaygani, "Why should we have a periodic safety and performance program for medical devices.," Biomedical Physics & Engineering, pp. 251-256, 2019.
[2] U.S. Food and Drug Administration. Infusion Pump Risk Reduction Strategies for Clinicians: Guidance for Industry and FDA Staff. Silver Spring, MD: FDA; 2014.
[3] International Electrotechnical Commission. IEC 60601-2-24:2012 — Medical Electrical Equipment — Part 2-24: Particular Requirements for the Basic Safety and Essential Performance of Infusion Pumps and Controllers. Geneva: IEC; 2012.
[4] European Parliament. Regulation (EU) 2017/745 of the European Parliament and of the Council on Medical Devices. Off J Eur Union. 2017;L117:1–175.
[5] Marcelle Di Angelis Ambar Felipe, Jos M. Latour, Maria Angélica Sorgini Peterlini, Mavilde Da Luz Gonçalves Pedreira. Placement of syringe infusion pumps and solution density can impact infusion performance: An experimental study. Journal of Neonatal Nursing, Volume 26, Issue 3, 2020.
[6] Tsao AC, Lovich JM, Parker MJ, Kost GJ, Bhaskaran A. Delivery interaction between co-infused medications: an in vitro study of microinfusion using syringe pumps. Anesth Analg. 2013;117(4):850–858.
[7]  Lannoy D, Décaudin B, Dewulf S, Simon N, Barthelemy C, Debaene B, Odou P. Infusion line setups including a syringe pump and a microbore infusion line reduce the risk of flow rate inaccuracies. Pharm World Sci. 2010;32(3):321–328.
[8] Claudia Saez. Impact of syringe size on the performance of infusion pumps at low flow rates. Pediatric Critical Care Medicine. 2010; 10(6):595-599.
[9] Hong KY, Lee EK, Kim Y, Choi DC, Min JJ. Effects of infusion tubing line lengths and syringe sizes on infusion system compliance: an experimental study using a syringe-type infusion pump at low flow rate. J Clin Monit Comput. 2023; 37(5):1379-1386.